CLEAN ROOM / PHARMA INDICATORS

UT-1408

  • TEMP + RH CONT. WITH SENSOR-LOW COST

     

    ♦ USFDA 21 CFR Part 11 - GAMP-4 Compatible

    ♦ Isolated 2 X 4-20 mA DC Re-Transmission Output For Temp & RH.

    ♦ Micro Controller Based Temp. + Rh Indicator Controller In Single Enclosure.

    ♦ Highly Stable Temp.+ Rh Sensor Available With Different Mounting.

    ♦ Potential Free Relay Contact Output For Temp. And Rh Up To 4 Nos.

    ♦ Process Parameter High And Low LED Status Indication For Both Channels.

    ♦ MODBUS RTU Mul Drop RS-485 Serial Communication For PLC, SCADA Etc

    ♦ Password Protection For Calibration And Programming Mode.

    ♦ Software Base Calibraon For Both Channels.

    ♦ User Programmable Offset Correction.

    ♦ Sensor Accuracy : RH - 2% / 3% , Temp.- 025 % Of Span.

    ♦  Ex. Flameproof Model Available For Hazardous Applications.

    ♦ Wall Mounting/FLP And Duct Mounted Sensor Available.

  •  FEATURE

     

    ♦ FDA 21 CFR part 11 - GAMP-4 compatible.

    ♦ Micro controller based Temp.+Rh indicator controller in single enclosure.

    ♦ All channel are programmable ranges and unit (Through Front key pad).

    ♦ Highly stable Temp.+ Rh sensor Available with different mounting.

    ♦ Potential Free Relay contact output For Temp. and Rh() up to 4 nos.

    ♦ Process parameter High and Low LEDs status indication for Both channels.

    ♦ MODBUS RTU multi drop RS-485 serial communication for PLC, SCADA etc.

    ♦ Password protection for calibration and programming mode.

    ♦ Software base Calibration for both channels.

    ♦ User programmable offset correction.

    ♦ Wall mounting/FLp and Duct mounted Sensor available.

    ♦ Isolated 2 X 4-20mA Retransmission output for Temp & Rh.

    ♦ Transmitter loop power supply with current limit as standard.

    ♦ Power supply :230V/110V AC. Or 24V/12V DC

    ♦ Sensor Accuracy : RH - 2% / 3% , Temp.- 025 % of Span.

    ♦ Ex. Flameproof model available for Hazardous applications.

INPUT Nos. Of Input 2 Nos. (Temp. + Rh)
Temperature Input External:4-20 mA Or 0-10 V / RTD Pt-100)
Humidity Input External: 0.8 To 3.9 V DC /0-5 V Dc /4-20 mA Or 0-10 V DC)
Transmitter Power Supply 24 V DC @ 30 mA Max, Isolated From I/P & O/P
Range Limits Temp :-40 To 100 *C & Humidity : 0-100 %
Accuracy Temp :0.25 % Fs & Humidity : 2% On Fs
Input Impedance >330 K For Voltage ,mV, And T/C, 100 R For Current
INDICATION Indication Temperature : 4 Digit And Humidity - 4 Digit - Red Led 7 Segment Display.
Status Indication Leds For Relay Status & Communication
Alarm Modes Very Low/Low/High/Very High
CONTROL MODE On Off Control Relay / SSR
Dead Band Controls High And Low Set Point For Level Application
OUTPUT Relay Output NO - C - NC Contacts , 5 A @240 V AC Resistive
SSR Output 1 / 2 / 3 / 4 Relays / SSR Output
Transmitter Power 24 V DC @ 20 mA ,12 V DC @ 20 mA
Retransmission Current & Voltage Outputs 2 Nos. 0~20/ 4~20 mA, 0 ~ 5 V , 0 ~ 1 0 V , 1 ~ 5V DC, Fully 3 Port Isolated
COMMUNICATION Communication Protocol & Method Modbus RTU RS- 485/RS- 232 Communication
Programmable Parameter & Max No. Baud Rate, Parity , Node Address , Stop Bit & 0-256 Unit
Wireless Communication GSM/GPRS Base Mobile(SMS) Communication
PROGRAMMBALE PARAMETERS Set point / Hysteresis / High Scale / Low Scale Full Range As Per Input Selection.
Relay Logic F.S. High & Low Alarm, Dead Band Alarm
Display Resolution -1999 To 9999 Fully Configurable For Linear Input
Programming Of Parameters Through 4 Tactile Keys
OTHER Dimensions & Mounting As Per Box Selection & Wall/Structure 2/2 Holes
(A) Supply Voltage (B) Power Consumption (A) 110/230 V AC/24 V DC, 50/60 Hz (B) 4 Watts Maximum
Operating Ambient Temperature And Humidity 0~50 °C , And Humidity : Below 90%, Non Condensing